-==== Common errors ====+===== Common errors ====
 + 
 +   * **IBG discharge failure.** If the ion beam gun power source shows the HLP26 or HLP27 error, reset the power source by pressing the button the says ''Discharge Enable / Standby'' with the label ''Reset'' next to it. A common cause for a discharge error is insufficient Ar pressure. A pressure of 3.8e-4 mbar is the lower boundary for a stable emission. At 3.7e-4 mbar to 3.6e-4 mbar there are insufficient Ar atoms for the IBG to maintain a stable discharge. An Ar flow rate of 3.0 sccm is sufficient, 2.5 sccm is on the lower end and might not be enough. **Note:** Pressure typically drops by 0.1e-4 mbar to 0.2e-4 mbar during etching over a scale of 1 to 3 minutes. Keep this in mind when selecting flow rate. 
  
-   * Discharge failure? If the power source shows the HLP26 or HLP27 error, reset the power source by pressing the button the says ''Discharge Enable / Standby'' 
  
  
e-beam_evaporator.1764755700.txt.gz · Last modified: 2025/12/03 09:55 by fabian