From Felicia@cryogenic.co.uk Tue Aug 29 16:38:23 2006
Received: from smtp001.apm-internet.net ([85.119.248.9]) by mail1.Physics.LeidenUniv.nl with Microsoft SMTPSVC(6.0.3790.1830);
	 Tue, 29 Aug 2006 16:38:47 +0200
Received: (qmail 79608 invoked from network); 29 Aug 2006 14:38:23 -0000
Received: from unknown (HELO cryogenic.co.uk) (213.120.114.238)
  by smtp001.apm-internet.net with SMTP; 29 Aug 2006 14:38:23 -0000
Content-class: urn:content-classes:message
MIME-Version: 1.0
Content-Type: text/plain;
  charset="iso-8859-1"
Content-Transfer-Encoding: quoted-printable
Subject: FW: problems with SMS120C Q5390
X-MimeOLE: Produced By Microsoft Exchange V6.5
Date: Tue, 29 Aug 2006 15:38:23 +0100
Message-ID: <3FBC6B0F963CBF41AE784D72A2D88C1A0F9828@CRYSRV01.cryogenic.local>
X-MS-Has-Attach: 
X-MS-TNEF-Correlator: 
Thread-Topic: problems with SMS120C Q5390
Thread-Index: AcbKwoi9upsjZK9VQBqUFOUqyBPT9wAr9szwAAChCgA=
From: "Felicia Doughty" <Felicia@cryogenic.co.uk>
To: <galli@physics.leidenuniv.nl>
Return-Path: Felicia@cryogenic.co.uk
X-OriginalArrivalTime: 29 Aug 2006 14:38:47.0851 (UTC) FILETIME=[D64CCFB0:01C6CB78]
Status: R
X-Status: NC
X-KMail-EncryptionState:  
X-KMail-SignatureState:  
X-KMail-MDN-Sent:  

=20
Dear Federica,

Repair will cost between =A3250 and =A3350.
To replace the linear power supply, it would be an extra =A3150.

If you chose to have it modified, this would include replacing =
conductors with FET reversing switch, an additional =A3900.

If you would like to go ahead with the repair or modification, please =
let us know and we will give you the appropriate authorization to send =
the Power Supply to us.

Best Regards,

Felicia Doughty,
Sales Administrator.
=20
Cryogenic Ltd
Tel: +44 20 87436049
Fax: +44 20 8749 5315
www.cryogenic.co.uk

-----Original Message-----
From: Mirek Owczarkowski=20
Sent: 29 August 2006 14:53
To: Felicia Doughty
Subject: FW: problems with SMS120C

=20

-----Original Message-----
From: Federica Galli [mailto:galli@physics.leidenuniv.nl]
Sent: 28 August 2006 17:59
To: Mirek Owczarkowski
Subject: problems with SMS120C

Dear Mirek,
after a few times we had again problems with our SMS120C power supply. =
This sort of problems (which I will describe below) we had already had =
occasionally before, but now it happens so frequently that normal magnet =
operation is hindered.

Problem 1: as I told you by phone last week, a few times after switching =
on the mains the error is given: "Power fail interrupt fault".=20

Problem 2: when ramping the field (current) to a given set point, both =
the actual output current/voltage and what is indicated on the front =
display start to go wrong. For example, when "ramping to 30A at the rate =
of 0.1A/s", after a while suddenly the voltage becomes negative on the =
display, the current decreases (both on display and on the output) often =
at a much higher rate than the preset, but still the given message is =
"ramping to 30A at the rate of 0.1A/s". If I kick gently on the display, =
suddenly the current increases again as it should and the voltage =
becomes positive. Occasionally the "wrong" behaviour occurs again, and =
again by tapping on the display it can be fixed for a short time. As I =
just said, it's not just a display
problem: when the display shows the negative values, the output reflects =
the values exactly. It has also occurred when ramping down the current; =
then suddenly the power supply would start to ramp up (positive =
voltage).

I have checked for loose connectors inside, in the vicinity of the =
display and on the boards themself, but found nothing obvious. I have =
the feeling that there is somewhere a loose contact on one of the =
boards, or a small crack. I also think that the 2 large relay, when =
switching, shake some connections around and some might have become =
progressively loose.

I tried to call up Cryogenic today, but I guess it's a bank holiday over =
there. Could you please send me a quotation as soon as possible, with =
mentioned the price for the repair and also the cost for the upgrade for =
the reverse switch? Could you also change the "old" low voltage power =
supply as you suggested by phone (the one inside the brass grid-case). =
And finally, would it be possible to receive from you a wooden crate for =
delivering the power supply in a safe way?


Thank you very much for your help and cooperation,

best regards,

Federica