From Roger@cryogenic.co.uk Thu Dec 21 14:59:57 2006
Received: from smtp001.apm-internet.net ([85.119.248.9]) by mail1.Physics.LeidenUniv.nl with Microsoft SMTPSVC(6.0.3790.1830);
	 Thu, 21 Dec 2006 14:59:34 +0100
Received: (qmail 62469 invoked from network); 21 Dec 2006 13:59:32 -0000
Received: from unknown (HELO cryogenic.co.uk) (213.120.114.238)
  by smtp001.apm-internet.net with SMTP; 21 Dec 2006 13:59:31 -0000
MIME-Version: 1.0
Content-Type: text/plain;
  charset="us-ascii"
Content-Transfer-Encoding: quoted-printable
Content-class: urn:content-classes:message
X-MimeOLE: Produced By Microsoft Exchange V6.5
Subject: RE: problem with helium probe 105
Date: Thu, 21 Dec 2006 13:59:57 -0000
Message-ID: <3FBC6B0F963CBF41AE784D72A2D88C1A16C3C0@CRYSRV01.cryogenic.local>
X-MS-Has-Attach: 
X-MS-TNEF-Correlator: 
Thread-Topic: problem with helium probe 105
Thread-Index: Acck9cXxtCG/DpIyStmgIXbWUkO3mQADlzXg
From: "Roger Mitchell" <Roger@cryogenic.co.uk>
To: "Federica Galli" <galli@physics.leidenuniv.nl>
Return-Path: Roger@cryogenic.co.uk
X-OriginalArrivalTime: 21 Dec 2006 13:59:34.0138 (UTC) FILETIME=[3E7835A0:01C72508]
Status: R
X-Status: NC
X-KMail-EncryptionState:  
X-KMail-SignatureState:  
X-KMail-MDN-Sent:  

Dear Federica,

The connections are as follows:

E =3D -V
D =3D +V
F=3D +I
C=3D -I

Where D & F are connected to the base of the superconducting wire=20
C is the connection in series with the heater which then connects to the
top of the s/c wire=20
E is the connection directly in series with the top of the s/c wire.

I regret to say the measurements do confirm the probe is damaged beyond
repair.  I will ask our sales department to send you a quotation for a
replacement. =20

Our records show that the 2.1mm stainless tube length should be 1253mm
long.  Please confirm this is correct if you order a replacement.  You
can withdraw the probe without problem from the cryostat

Best regards & Merry Christmas,

Roger Mitchell



-----Original Message-----
From: Federica Galli [mailto:galli@physics.leidenuniv.nl]=20
Sent: 21 December 2006 11:53
To: Roger Mitchell
Cc: Kelly Simon
Subject: problem with helium probe 105

Dear Roger,
I just phoned this morning about an issue with a helium probe:

here are the details:

1. probe type:105
2. controle unit: HLG200/210
3. stainless steel cryostat project nr: J2024 4. magnet insert project
nr: J2041

Fault findings:=20

1. Level indication does not make sense (empty=3D560mm, full=3D810mm,
insetad of empty=3D0mm, full=3D555mm).
2. Resistance between E-C=3D5 Ohm, OK
3. Resistance between E-F=3D1 Ohm, WRONG according to manual (should be
95-105
Ohm)
4. Resistance between F-D=3DO.L. (open), WRONG according to manual =
(should
be
0.5-4 Ohm)

The resistances have been measured at the probe connector.

Could you please tell me what the resistances/connections above mean?

Could you also send me a quotation for a new probe?

Best regards,