 ====== Manual ====== ====== Manual ======
 +===== Start up & Loading sample =====
 +  * Open N2 line at the wall 1/2 turn (check pressure reducer)
 +  * Open Ar bottle and the in-line valve in the yellow cabinet
 +  * Check if the required target material is in 
 +  * If system is vented, insert your sample and close the top
 +  * Switch on system mains (green switch)
 +  * Switch on chiller (below the table)
 +
 ===== Sputtering ===== ===== Sputtering =====
-  * TODO+  * Purge the argon line by opening the needle valve and repeatedly opening and closing the argon valve at the bottle 
 +  * Carefully close the needle valve to pump down 
 +  * Switch turbo on at 1e-1 mbar by pressing "pumping unit" and "stand by" on the turbo controller  
 +  * Adjust pressure using needle valve to 5E-2mbar 
 +  * Set current dial knob to 0 and switch to "HTS" 
 +  * Select sputtering time and press start 
 +  * Adjust current dial for desired current (lower-righ indicator!) 
 +  * Wait 
 +  * Switch turbo pump off 
 +  * If turbo stands still, switch off machine (green button), automatic N2 purging starts 
 +  * Close Ar, N2 supply ; switch off chiller 
 + 
 ===== change target ===== ===== change target =====
   * open sputter head, fix head,   * open sputter head, fix head,
Line 17: Line 37:
  
  
-^ Material              Date    ^ Sample ID   ^ Process parameters                  ^ Measurement ^ Result      ^ Xtal    ^ Rate          ^ +^ Material              Date    ^ purpose        ^ Process parameters    ^ Measurement ^ Result      ^ Rate          ^ 
-XXXX                   20060913 13_9_6_5    4.0e-mbar, 200 mA, 5 min, tilted  X-ray       110 nm      46.8 nm | 22 nm/min     |+Ag (?)               20110623 SEM of SU-8    5E-mbar, 15 mA, 15s SEM pic     nice              |
  
 +====== bugs ======
 +===== yellow HV light does not go on =====
 +Don't do this because of high voltage in the machine!
  
 +{{:balzer-med010-relay.jpg?700|}}