| ====== Manual ====== | ====== Manual ====== | ||
| + | ===== Start up & Loading sample ===== | ||
| + | * Open N2 line at the wall 1/2 turn (check pressure reducer) | ||
| + | * Open Ar bottle and the in-line valve in the yellow cabinet | ||
| + | * Check if the required target material is in | ||
| + | * If system is vented, insert your sample and close the top | ||
| + | * Switch on system mains (green switch) | ||
| + | * Switch on chiller (below the table) | ||
| + | |||
| ===== Sputtering ===== | ===== Sputtering ===== | ||
| - | * TODO | + | * Purge the argon line by opening the needle valve and repeatedly opening and closing the argon valve at the bottle |
| + | * Carefully close the needle valve to pump down | ||
| + | * Switch turbo on at 1e-1 mbar by pressing " | ||
| + | * Adjust pressure using needle valve to 5E-2mbar | ||
| + | * Set current dial knob to 0 and switch to " | ||
| + | * Select sputtering time and press start | ||
| + | * Adjust current dial for desired current (lower-righ indicator!) | ||
| + | * Wait | ||
| + | * Switch turbo pump off | ||
| + | * If turbo stands still, switch off machine (green button), automatic N2 purging starts | ||
| + | * Close Ar, N2 supply ; switch off chiller | ||
| + | |||
| ===== change target ===== | ===== change target ===== | ||
| * open sputter head, fix head, | * open sputter head, fix head, | ||
| Line 17: | Line 37: | ||
| - | ^ Material | + | ^ Material |
| - | | XXXX | 20060913 | + | | Ag (?) | 20110623 |
| + | ====== bugs ====== | ||
| + | ===== yellow HV light does not go on ===== | ||
| + | Don't do this because of high voltage in the machine! | ||
| + | {{: | ||