electron_beam_lithography
Differences
This shows you the differences between two versions of the page.
| Both sides previous revisionPrevious revisionNext revision | Previous revision | ||
| electron_beam_lithography [2008/05/20 07:42] – edo | electron_beam_lithography [2008/06/03 10:54] (current) – verstoep | ||
|---|---|---|---|
| Line 31: | Line 31: | ||
| **In a lithographic step, try to avoid changes of the write field size that require a change in the working distance, because then recalibration of writing field and UV system are needed.** | **In a lithographic step, try to avoid changes of the write field size that require a change in the working distance, because then recalibration of writing field and UV system are needed.** | ||
| + | |||
| Line 36: | Line 37: | ||
| To rotate a group of structures: | To rotate a group of structures: | ||
| - | - select | + | - select |
| - go to menu modify | - go to menu modify | ||
| - select rotate free | - select rotate free | ||
| Line 44: | Line 45: | ||
| =====Beam current===== | =====Beam current===== | ||
| When the layers and working areas of your sample have been designed, it is time to think about the beam current. See the calibration plot at the machine for spot sizes. Use 20-100 pA for small features and 5 nA for large ones as a start, make adjustments if this leads to excessive writing times (the higher the current the larger the spot, the smaller the spot the smaller the current and therefore the larger the exposure time fore a given pattern). | When the layers and working areas of your sample have been designed, it is time to think about the beam current. See the calibration plot at the machine for spot sizes. Use 20-100 pA for small features and 5 nA for large ones as a start, make adjustments if this leads to excessive writing times (the higher the current the larger the spot, the smaller the spot the smaller the current and therefore the larger the exposure time fore a given pattern). | ||
| + | |||
| Line 61: | Line 63: | ||
| - Press Calculator | - Press Calculator | ||
| - Take a look at the Beam Current VS Spot Size graph, which is on the cover of the manual of the SEM which you will find on the console. | - Take a look at the Beam Current VS Spot Size graph, which is on the cover of the manual of the SEM which you will find on the console. | ||
| - | - If you want to have a 100 nm resolution (80 nm is about the best of this system), you will need 10< | + | - If you want to have a 100 nm resolution |
| - Set your Area Dose to the value required by your process. (For hardening resist, you typically require between 40 and 160 uAs/ | - Set your Area Dose to the value required by your process. (For hardening resist, you typically require between 40 and 160 uAs/ | ||
| - | - Use the calculate Icon to calculate the missing parameter(s). | + | - Use the calculate Icon to calculate the missing parameter(s) |
| + | - the button [Times] shows you how long the etching will take. | ||
| ====== Start-Up ======= | ====== Start-Up ======= | ||
| Line 72: | Line 76: | ||
| - Load the sample into the SEM by using the loadlock which is controlled by the red button. With this single button you can vent and pump down the loadlock. The button lights up if the loadlock is venting or pumping down. Read the [[JEOL 820 SEM# | - Load the sample into the SEM by using the loadlock which is controlled by the red button. With this single button you can vent and pump down the loadlock. The button lights up if the loadlock is venting or pumping down. Read the [[JEOL 820 SEM# | ||
| - Slide the sample holder inside the vacuum chamber onto the stage. | - Slide the sample holder inside the vacuum chamber onto the stage. | ||
| + | |||
| + | |||
| ===== System Start-Up ===== | ===== System Start-Up ===== | ||
| - Turn on the display of the computer. Log in to Elphy Quantum. | - Turn on the display of the computer. Log in to Elphy Quantum. | ||
| - | - Set the working distance of the SEM to 10 mm. (OR: 8 for exceptional situations, 3x3 or 4x4 marker for large fields) | + | - Set the working distance of the SEM to 10 mm (8 in exceptional situations) for small structures, 3x3 for a 3x3 mm working field or 4x4 marker for a 4x4 mm working field |
| - check that aperture (Diafragma) ' | - check that aperture (Diafragma) ' | ||
| - Drive the stage to the chessy position with the '' | - Drive the stage to the chessy position with the '' | ||
| + | - Check that PCD is on. | ||
| - Check that the voltage is set to 20 kV and that the filament current is 0 A. | - Check that the voltage is set to 20 kV and that the filament current is 0 A. | ||
| - Turn on the main power (red square button). It will light up. | - Turn on the main power (red square button). It will light up. | ||
| Line 95: | Line 102: | ||
| - Set the current to about 1 nA (With bottom right knobs) | - Set the current to about 1 nA (With bottom right knobs) | ||
| - Improving the image | - Improving the image | ||
| - | - Use the knobs on the bottom console to change brightness and contrast. The option coarse allows you to toggle between making large or small adjustments. | + | - Use the knobs on the bottom console to change brightness and contrast. |
| - Use magnification and the position joystick to change the position you look at. | - Use magnification and the position joystick to change the position you look at. | ||
| - | - Focus the image (you should be able to have a 100 nm resolution) | + | - Focus the image (you should be able to have a 100 nm resolution). The option coarse allows you to toggle between making large or small adjustments. |
| - Change the current with the bottom right knobs '' | - Change the current with the bottom right knobs '' | ||
| - iterate the above steps. | - iterate the above steps. | ||
| Line 106: | Line 113: | ||
| ====== Calibration ====== | ====== Calibration ====== | ||
| + | |||
| ===== Calibrate the Write field on the chessy ===== | ===== Calibrate the Write field on the chessy ===== | ||
| Line 114: | Line 122: | ||
| - Go to desktop 3 | - Go to desktop 3 | ||
| - | - Set Magnification to desired value (for example 100X) | + | |
| - | - Drive the stage to the center of a chessy block (point at it and click ''< | + | |
| - Set the hardware to: slow 2 / EBPG / Beam Blanking external | - Set the hardware to: slow 2 / EBPG / Beam Blanking external | ||
| - | - Select your writing field (you can use 400 (@ 50X ?) um OR 800 um (@ 100X)) | + | - Select your writing field in microscope control and click '' |
| + | - Set Magnification with the knob to the corresponding value you have entered for the writing field. | ||
| - Open a new image (file-> | - Open a new image (file-> | ||
| - The image should have the right size and not be rotated (it doesn' | - The image should have the right size and not be rotated (it doesn' | ||
| - **Course: | - **Course: | ||
| - | - Drive the stage to the center of a chessy block (point at it and click ''< | + | - Drive the stage to the center of a chessy block (point at it and click ''< |
| - | - Open a writefield calibration positionlist (file-> | + | - Open a writefield calibration positionlist (file-> |
| - Scan the image. Maximize the window of the image. | - Scan the image. Maximize the window of the image. | ||
| - Markers appear in the image | - Markers appear in the image | ||
| - Drag (ctrl-leftclick) all 4 markers to the right place. They have to be moved! | - Drag (ctrl-leftclick) all 4 markers to the right place. They have to be moved! | ||
| - | - To apply the alignment press '' | + | - To apply the alignment press '' |
| - **Fine:** | - **Fine:** | ||
| - Optionally scan the image again. | - Optionally scan the image again. | ||
| Line 135: | Line 143: | ||
| - Iterate the above steps if necessary. | - Iterate the above steps if necessary. | ||
| - Close the image window (Don't save), Close the postionlist (Don't save) | - Close the image window (Don't save), Close the postionlist (Don't save) | ||
| - | - Set '' | + | - Set '' |
| - | - Press send (Where?) | + | - Press send in the '' |
| ===== UV coordinate system calibration ===== | ===== UV coordinate system calibration ===== | ||
| Line 171: | Line 179: | ||
| ====== Writing ====== | ====== Writing ====== | ||
| + | |||
| ===== Preparing your program ===== | ===== Preparing your program ===== | ||
| Line 177: | Line 186: | ||
| - Load the positionlist you want to write: file-> | - Load the positionlist you want to write: file-> | ||
| - Make sure your Area Dose corresponds to your process (For hardening resist, you typically require between 40 and 160 uAs/ | - Make sure your Area Dose corresponds to your process (For hardening resist, you typically require between 40 and 160 uAs/ | ||
| - | - Make sure you have selected the right layer for exposure. | + | - Make sure you have selected the right layer for exposure. |
| - | | + | |
| There are two ways to do the etching: with beam stabilizer OR without beam stabilizer. | There are two ways to do the etching: with beam stabilizer OR without beam stabilizer. | ||
electron_beam_lithography.1211269342.txt.gz · Last modified: 2008/05/20 07:42 by edo