User Tools

Site Tools


e-beam_evaporator

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revisionPrevious revision
Next revision
Previous revision
e-beam_evaporator [2025/06/13 13:54] – [Aftermath] wigboute-beam_evaporator [2025/12/03 16:13] (current) – [Common errors] fabian
Line 182: Line 182:
    * Open the Ar-valve.    * Open the Ar-valve.
    * Wait until the chamber pressure is stable.    * Wait until the chamber pressure is stable.
-   * Set your desired Argon flow (e.g. 2.sccm)+   * Set your desired Argon flow (e.g. 3.sccm)
    * Check all the settings from the ion beam (voltages, etc.)    * Check all the settings from the ion beam (voltages, etc.)
    * Press ''discharge''    * Press ''discharge''
Line 195: Line 195:
    * Close Ar-valve.    * Close Ar-valve.
    * Put back the sampleholder to the original position, facing downwards.    * Put back the sampleholder to the original position, facing downwards.
 +
 +=== Etch rate ===
 +
 +The etch rate in theory is about half of the etch speed of the ion beam etcher in the clean room. Beam diameter of the clean room IBG is approx. 4 cm, beam diameter of the e-beam IBG is a little below 10 cm. Beam current of the clean room IBG is ~7.5 mA, beam current in the e-beam IBG is ~15 mA.
 +
 +The intensity is given as beam current divided by beam area (I/r^2π) where r is the radius of the beam (half the beam diameter).
 +
 +In practice, however, the etch rate is about 1/3 the etch rate of the clean room ion beam etcher (tested 02/12/2025 with 60 nm of Nb, capped with 5 nm Pt).
  
 ---- ----
Line 251: Line 259:
  
  
-==== Common errors ====+===== Common errors ====
 + 
 +   * **IBG discharge failure.** If the ion beam gun power source shows the HLP26 or HLP27 error, reset the power source by pressing the button the says ''Discharge Enable / Standby'' with the label ''Reset'' next to it. A common cause for a discharge error is insufficient Ar pressure. A pressure of 3.8e-4 mbar is the lower boundary for a stable emission. At 3.7e-4 mbar to 3.6e-4 mbar there are insufficient Ar atoms for the IBG to maintain a stable discharge. An Ar flow rate of 3.0 sccm is sufficient, 2.5 sccm is on the lower end and might not be enough. **Note:** Pressure typically drops by 0.1e-4 mbar to 0.2e-4 mbar during etching over a scale of 1 to 3 minutes. Keep this in mind when selecting flow rate. 
  
-   * Discharge failure? If the power source shows the HLP26 or HLP27 error, reset the power source by pressing the button the says ''Discharge Enable / Standby'' 
  
  
e-beam_evaporator.1749822895.txt.gz · Last modified: 2025/06/13 13:54 by wigbout

Donate Powered by PHP Valid HTML5 Valid CSS Driven by DokuWiki