User Tools

Site Tools


e-beam_evaporator

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revisionPrevious revision
e-beam_evaporator [2025/12/03 16:12] – [Common errors] fabiane-beam_evaporator [2025/12/03 16:13] (current) – [Common errors] fabian
Line 259: Line 259:
  
  
-==== Common errors ====+===== Common errors =====
  
-   Discharge failureIf the ion beam gun power source shows the HLP26 or HLP27 error, reset the power source by pressing the button the says ''Discharge Enable / Standby'' with the label ''Reset'' next to it. A common cause for a discharge error is insufficient Ar pressure. A pressure of 3.8e-4 mbar is the lower boundary for a stable emission. At 3.7e-4 mbar to 3.6e-4 mbar there are insufficient Ar atoms for the IBG to maintain a stable discharge. An Ar flow rate of 3.0 sccm is sufficient, 2.5 sccm is on the lower end and might not be enough. **Note:** Pressure typically drops by 0.1e-4 mbar to 0.2e-4 mbar during etching over a scale of 1 to 3 minutes. Keep this in mind when selecting flow rate.+   **IBG discharge failure.** If the ion beam gun power source shows the HLP26 or HLP27 error, reset the power source by pressing the button the says ''Discharge Enable / Standby'' with the label ''Reset'' next to it. A common cause for a discharge error is insufficient Ar pressure. A pressure of 3.8e-4 mbar is the lower boundary for a stable emission. At 3.7e-4 mbar to 3.6e-4 mbar there are insufficient Ar atoms for the IBG to maintain a stable discharge. An Ar flow rate of 3.0 sccm is sufficient, 2.5 sccm is on the lower end and might not be enough. **Note:** Pressure typically drops by 0.1e-4 mbar to 0.2e-4 mbar during etching over a scale of 1 to 3 minutes. Keep this in mind when selecting flow rate.
  
  
e-beam_evaporator.txt · Last modified: 2025/12/03 16:13 by fabian

Donate Powered by PHP Valid HTML5 Valid CSS Driven by DokuWiki